Model 2024 2% Ammonia (NH3)
The VALTRONICS Model 2024-2% NH3 is a non-dispersive infrared, gas diffusion type, gas sampling NH3 monitor for use as an OEM module. It produces a control signal proportional to NH3 concentration. The linear signal outputs of 0-5 VDC and 4-20 mA current loop may be used with any central control alarm of air monitoring system.
Availability: Commercially Available
Manufacturer |
| VTI-Valtronics, Inc. |
Product Name |
| Model 2024 2% Ammonia (NH3) |
Detection Category |
| Chemical |
Detection Principle |
| Molecular Spectroscopy; |
Detection Method |
| Non-Dispersive Infrared (NDIR); |
Application |
| Gas Analysis; |
Equipment Type |
| Sensor |
Product Synopsis |
| The VALTRONICS Model 2024-2% NH3 is a non-dispersive infrared, gas diffusion type, gas sampling NH3 monitor for use as an OEM module. It produces a control signal proportional to NH3 concentration. The linear signal outputs of 0-5 VDC and 4-20 mA current loop may be used with any central control alarm of air monitoring system. |
Availability |
| Commercially Available |
Technology Readiness Level (TRL) |
| 9 |
User Feedback Sources |
| Indifferent/No user feedback |
Dimensions |
| 7.75 x 4.5 x 3.75 in (197 x 115 x 96 mm) |
Weight |
| <1 lb.; < 0.5 lbs (.23 kg) |
Power Requirements |
| AC/DC Line Power; External power: 14.0-24.0 VDC absolute min/max |
Transportability |
| Pocket Size |
Ruggedness |
| Waterproof/Resistant |
Operating Conditions |
| 5 to 40°C (operating temperature) -40 to 70°C (storage temperature) 0 – 95% Relative Humidity, non-condensing |
Consumables |
| <$1,000/year; Nitrogen and carbon dioxide calibration gases, regulator, flow meter and tubing |
Calibration Schedule |
| Yes, zero calibration must be done every 6 months |
Suggested Routine Maintenance |
| Rarely |
Data Analysis Support Equipment |
| OEM optical bench, precision gas calibration kit |
Shelf Life |
| >5 years |
Unit Cost |
| $1,650.00 |
Maintenance Cost |
| ~$200, infrequently needed |
Other Chemical Targets |
| Ammonia and refrigerant |
Sample Introduction |
| Vapor/Gas |
Sensitivity/Detection Limits |
| ppm; % |
Start Up Time (From Cold Start To Sample Ready) |
| 30 to 60 seconds |
Response Time (Sample Application To Output) |
| Continuous (<30 seconds); <10 seconds |
Alarm Capability |
| Yes |
Other Operational Parameters |
| $2,245 ammonia, $1,650 for refrigerant |
Training Required |
| <1 week training |
Manuals Available |
| User Manual |
Warranties |
| 1 yr |