MALDImini-1

The MALDImini-1 is a high-sensitivity MALDI MS capable of MSn measurements. It is nonetheless astonishingly compact thanks to the use of pioneering Digital Ion Trap (DIT) technology. It fits on a bench space equivalent to an A3 piece of paper

Notify me on updates of this product 

Availability: Commercially Available

Greg Feldman
Sr. Field Sales Engineer
gsfeldman@shimadzu.com
1-913-888-9449, Ext. 1270
8052 Reeder Street
Lenexa, KS 66214
Midwest Regional Office
Jason Kuchar, Ph. D.
Life Science account Manager
jakuchar@shimadzu.com
1-913-888-9449, Ext. 1265
8052 Reeder Street
Lenexa, KS 66214
Midwest Regional Office
www.ssi.shimadzu.com
Manufacturer
Shimadzu
Product Name
MALDImini-1
Detection Category
Biological
Detection Principle
MALDI-ion trap Mass Spectrometry;
Detection Method
MALDI-mass spectrometry;
Application
Detection;
Product Model Number
MALDImini-1
Equipment Type
Instrument
Product Synopsis
The MALDImini-1 is a high-sensitivity MALDI MS capable of MSn measurements. It is nonetheless astonishingly compact thanks to the use of pioneering Digital Ion Trap (DIT) technology. It fits on a bench space equivalent to an A3 piece of paper
Availability
Commercially Available
Technology Readiness Level (TRL)
9
Market Entry Date
2019
User Feedback Sources
Indifferent/No user feedback
Dimensions
10.5 x 15.2 x 12.6 in (26.8 x 38.5 x 32.0 cm)
Weight
>50 Lbs.; 55 Lbs. (25 kg)
Power Requirements
AC 100 to 240 V, 50/60 Hz, 960 VA
Noise Produced
Negligible
Transportability
Benchtop/Fixed; Small Footprint
Operating Conditions
18 to 26 °C; 40 to 70% relative humidity (non-condensing)
Solvents/Reagents
MALDI matrix
Calibration Schedule
Two-point calibration with analysis
Suggested Routine Maintenance
Minimal
Available Accessories
Target Plates
Data Analysis Support Equipment
Requires third-party data processing software
Communications Interface
Desktop Computer
Shelf Life
>5 years
Unit Cost
$250,000.00
Maintenance Cost
Minimal
Other Biological Targets
Proteins, peptides, glycans
Sample Introduction
Liquid
Sensitivity/Detection Limits
10 fmol Glu-Fib, 500 fmol BSA Mass Range: 650-70000 m/z Mass Accuracy: <200 ppm (external standard, m/z 1000 to 5000)
Start Up Time (From Cold Start To Sample Ready)
<15 minutes
Response Time (Sample Application To Output)
<5 minutes
Software Control
On-board
Other Operational Parameters
MSn capable
Training Required
<1 day training
Training Available
Yes
Manuals Available
Yes
Reachback Service
Yes
IATA Shipping Restrictions
Yes; Argon and Helium requirements
Approved Vendors
Shimadzu
Data File Type Format
.mzML, .mzXML
Please enter contact info.
Email Address: 
Company: 
Name: 
Title: