PM1704 Series

The PM1704 series handheld radiation detectors combine radiation detection, radioisotope identifier, and dosimeter capabilities. The PM1704 is a basic model of the PM1704 product line containing a single, spectroscopic CsI detector with a 13 mR/h maximum exposure rate range (versus 1,300 R/h for the PM1704M). The PM1704A-M uses CsI and GM detectors with gamma sensitivity of no less than 1 cps/µrem/h.

Notify me on updates of this product 

Availability: Discontinued

James Mehalchick
info@polimaster.us
+1 703 525 5075
44873 Falcon Place Suite 128
Sterling, VA 20166
USA
www.polimaster.com
Manufacturer
Polimaster, Inc.
Product Name
PM1704 Series
Detection Category
Radiological/Nuclear;
Detection Principle
Scintillation Detectors;Geiger-Muller Detectors;
Detection Method
Scintillation Detectors;Geiger-Muller Detectors;
Application
Personal Radiation Detectors;
Product Model Number
PM1704A, PM1074A-M, PM1704, PM1704M
Equipment Type
Instrument
Product Synopsis
The PM1704 series handheld radiation detectors combine radiation detection, radioisotope identifier, and dosimeter capabilities. The PM1704 is a basic model of the PM1704 product line containing a single, spectroscopic CsI detector with a 13 mR/h maximum exposure rate range (versus 1,300 R/h for the PM1704M). The PM1704A-M uses CsI and GM detectors with gamma sensitivity of no less than 1 cps/µrem/h.
Availability
Discontinued
Technology Readiness Level (TRL)
9
User Feedback Sources
Indifferent/No user feedback
Dimensions
(A and A-M) 2.9 x 3.4 x 1.5 in (7.4 x 8.6 x 3.81 cm) (M and 1704) 5.2 x 2.3 x 1.9 (13.2 x 5.8 x 4.8 cm)
Weight
<1 lb.; 0.45 kg
Power Requirements
Alkaline Battery; AA battery (300 hours)
Transportability
Handheld
Ruggedness
Extreme Temperatures/Humidity; Waterproof/Resistant; Dust; IP65
Operating Conditions
-20 to 50 °C (- 4 to 122 °F), <98% RH
Consumables
None
Solvents/Reagents
None
Calibration Schedule
Self-check, and calibration lasts 6 months under routine use
Suggested Routine Maintenance
None
Available Accessories
Belt clip or web loop holster
Data Analysis Support Equipment
None
Communications Interface
Bluetooth, GPS
Expected Operational Life
Indefinite
Shelf Life
>5 years
Unit Cost
$3,000.00
Maintenance Cost
None
Radiological Targets
Gamma
Adaptability
No; cannot detect or add new targets
Sample Introduction
Radioactive Particle
Sensitivity/Detection Limits
µR
Response Time (Sample Application To Output)
Continuous (<30 seconds)
Total Run Time
Continuous monitor
Alarm Capability
Audible and visual
Software Control
On-board
Other Operational Parameters
Detector: CsI, Dose Rate: 13 mR/h (A, 1704), 1,300 R/H (A-M, M), Data storage: 1,000 spectra and 15,000 events (A, A-M) or 100 spectra nd 50,000 events (1704, M)
Training Required
<1 day training
Training Available
Yes
Manuals Available
Yes
Please enter contact info.
Email Address: 
Company: 
Name: 
Title: