RadTarge II Series

RadTarge II Series is designed to process routine personnel dosimetry on which occupational dose of record is based. This pager-like, direct-reading dosimeter accurately detects and measures radiation exposure for workers and responders in potentially hazardous environments. The upward-facing, backlit display offers users readability in dark or bright environments. The Mini and W models are designed for strong radiation fields, and the H model for weak.

Notify me on updates of this product 

Availability: Commercially Available

contact@x-zlab.com
+1 925 355 5199
231 Market Place Suite 728
San Ramon, CA 94583
USA
www.x-zlab.com
Manufacturer
X-Z Lab Inc.
Product Name
RadTarge II Series
Detection Category
Radiological/Nuclear
Detection Principle
Scintillation Detectors;
Detection Method
Scintillation Detectors;
Application
Personal Radiation Detectors;
Product Model Number
Mini, W, H
Equipment Type
Instrument
Product Synopsis
RadTarge II Series is designed to process routine personnel dosimetry on which occupational dose of record is based. This pager-like, direct-reading dosimeter accurately detects and measures radiation exposure for workers and responders in potentially hazardous environments. The upward-facing, backlit display offers users readability in dark or bright environments. The Mini and W models are designed for strong radiation fields, and the H model for weak.
Availability
Commercially Available
Technology Readiness Level (TRL)
9
User Feedback Sources
Indifferent/No user feedback
Dimensions
2.7 x 1.8 x 0.7 in (6.9 x 4.6 x 1.7 cm)
Weight
<1 lb.; 2.1 oz (60 g)
Power Requirements
Lithium Battery; Rechargeable Lithium ion battery (200 hours)
Noise Produced
80 dB
Transportability
Pocket Size
Ruggedness
Extreme Temperatures/Humidity; Waterproof/Resistant; Dust; IP65
Operating Conditions
-20 to 50 °C (- 4 to 122 °F)
Consumables
None
Solvents/Reagents
None
Suggested Routine Maintenance
None
Available Accessories
None
Data Analysis Support Equipment
RadSuite-Dose (Mac/PC software)
Communications Interface
MicroUSB
Expected Operational Life
Indefinite
Shelf Life
>5 years
Maintenance Cost
None
Radiological Targets
Gamma, X-ray
Adaptability
No; cannot detect or add new targets
Sample Introduction
Radioactive Particle
Sensitivity/Detection Limits
µSv
Response Time (Sample Application To Output)
Continuous (<30 seconds); < 8 seconds
Total Run Time
Continuous monitor
Alarm Capability
Audible, visual, vibrational
Software Control
On-board, PC
Other Operational Parameters
Detector: silicon photomultiplier and YSO scintillator, Dose Rate: 1 µrem/h–10 rem/h (Mini), 1 µrem/h–100 mrem/h (H), 10 µrem/h–10 rem/h (W), Energy Range: 30 keV–3 MeV
Training Required
<1 day training
Training Available
Yes
Manuals Available
Yes
Please enter contact info.