Monitor 1000 EC

The Monitor 1000EC is an energy compensated radiation detector that measures gamma, and x-rays. Users can choose from readings of CPM, CPS, µSv/hr, mR/hr, or in accumulated counts.

Notify me on updates of this product 

Availability: Commercially Available

raditioninfo@seintl.com
+1 800 293 5759
P.O. Box 39, 436 Farm Road
Summertown, TN 38483
USA
www.seintl.com
Manufacturer
S.E. International, Inc.
Product Name
Monitor 1000 EC
Detection Category
Radiological/Nuclear;
Detection Principle
Geiger-Muller Detectors;
Detection Method
Geiger-Muller Detectors;
Application
Personal Radiation Detectors;
Equipment Type
Instrument
Product Synopsis
The Monitor 1000EC is an energy compensated radiation detector that measures gamma, and x-rays. Users can choose from readings of CPM, CPS, µSv/hr, mR/hr, or in accumulated counts.
Availability
Commercially Available
Technology Readiness Level (TRL)
9
User Feedback Sources
Indifferent/No user feedback
Dimensions
5.5 x 2.7 x 1.3 in (14 x 6.8 x 3.3 cm)
Weight
<1 lb.; 0.45 lbs. (204g)
Power Requirements
Alkaline Battery; 2 AA alkaline batteries (500 hours)
Noise Produced
70 db at 1 m
Transportability
Handheld
Operating Conditions
-10 to 50°C (14 to 122°F)
Consumables
None
Solvents/Reagents
None
Calibration Schedule
NIST Calibration
Suggested Routine Maintenance
None
Available Accessories
Carrying Case, Xtreme Boot, Stand, Mini-USB Cable
Data Analysis Support Equipment
Observer Software for PCs and Android OS
Communications Interface
USB, Bluetooth 4.1
Expected Operational Life
Indefinite
Shelf Life
>5 years
Maintenance Cost
None
Radiological Targets
Gamma, X-ray
Adaptability
No; cannot detect or add new targets
Sample Introduction
Radioactive Particle
Sensitivity/Detection Limits
mR; µR; 0.001 (1µR) to 1000 mR/hr
False Positive Rates
±15%
False Negative Rates
±15%
Response Time (Sample Application To Output)
Continuous (<30 seconds)
Total Run Time
Continuous
Alarm Capability
Audible, visual
Software Control
On-board, PC
Other Operational Parameters
Users can choose from readings of CPM, CPS, µSv/hr, mR/hr, or in accumulated counts.
Training Required
<1 hour training
Training Available
Yes
Manuals Available
Yes
Warranties
1 year limited