700 Series
The portable Model 702i isotopic measurement system was developed to give end users such as first responders a simple tool to quickly locate any abnormal levels of radioactivity, and to accurately identify the isotopes present. The instrument houses an internal 5.1 x 3.8 cm (2 x 1.5 in.) NaI detector. The signal is gain stabilized, and the instrument automatically calibrates itself via an embedded 40K source. The 711i contains a LaBr detector and identifies mixed isotopes
Availability: Commercially Available
Manufacturer |
| Ludlum Measurements, Inc. |
Product Name |
| 700 Series |
Detection Category |
| Radiological/Nuclear; |
Detection Principle |
| Scintillation Detectors; |
Detection Method |
| Scintillation Detectors; |
Application |
| Identifiers; |
Product Model Number |
| 702i, 711i, 703e, 711e |
Equipment Type |
| Instrument |
Product Synopsis |
| The portable Model 702i isotopic measurement system was developed to give end users such as first responders a simple tool to quickly locate any abnormal levels of radioactivity, and to accurately identify the isotopes present. The instrument houses an internal 5.1 x 3.8 cm (2 x 1.5 in.) NaI detector. The signal is gain stabilized, and the instrument automatically calibrates itself via an embedded 40K source. The 711i contains a LaBr detector and identifies mixed isotopes |
Availability |
| Commercially Available |
Technology Readiness Level (TRL) |
| 9 |
User Feedback Sources |
| Indifferent/No user feedback |
Supporting Documentation |
| Docs: Ludlum_702i isotope identifier.pdf; |
Dimensions |
| 9.9 x 4.5 x 8.7 in (25.1 x 11.4 x 22.1 cm) |
Weight |
| <30 lbs.; 5.2 lbs. (2.4 kg) |
Power Requirements |
| NiMH Battery; Rechargeable NiMH AA batteries (8 hours) |
Transportability |
| Handheld |
Ruggedness |
| Extreme Temperatures/Humidity; Waterproof/Resistant; Dust; IP56 |
Operating Conditions |
| -20 to 50 °C (-4 to 122 °F), < 95% RH |
Consumables |
| None |
Solvents/Reagents |
| None |
Calibration Schedule |
| Automatic calibration (temperature) stabilization with low-level 40K source. Coarse and fine energy calibration and dose-rate calibration done at factory, but available for expert users. |
Suggested Routine Maintenance |
| None |
Available Accessories |
| AC power adapter, RS-232 adapter cable |
Data Analysis Support Equipment |
| 320 x 240 pixel, high brightness, 32,000-color, 8.9 cm (3.5 in.) transflective LCD display |
Communications Interface |
| 10/100 Ethernet port |
Expected Operational Life |
| Indefinite |
Shelf Life |
| >5 years |
Maintenance Cost |
| None |
Radiological Targets |
| Gamma |
Sample Introduction |
| Radioactive Particle |
Sensitivity/Detection Limits |
| µSv; cps; uR; 1292 cps per µSv/h (775 cpm per µR/hr) |
Response Time (Sample Application To Output) |
| Continuous (<30 seconds) |
Total Run Time |
| Continuous |
Alarm Capability |
| Audible, visual |
Software Control |
| On-board, Remote, PC |
Other Operational Parameters |
| Energy Range; 18 keV to 3 MeV, Energy Resolution: 7.5 to 8.5% |
Training Required |
| <1 day training |
Training Available |
| Yes |
Manuals Available |
| Yes |
Data File Type Format |
| ANSI N42.42 |