Rapid Response Fentanyl (FYL) Forensic Test Kit
The Rapid ResponseTM Fentanyl (FYL) Forensic Test Kit is a lateral flow chromatographic immunoassay for the qualitative detection of Fentanyl/Norfentanyl in liquid and powder substances at the cut-off concentration of 400 ng/mL. It is specific for Fentanyl screening with no significant cross reactivity to other opiates, such as Morphine and Heroin. The Rapid ResponseTM Fentanyl (FYL) Forensic Test Kit is able to detect Fentanyl and many other Fentanyl analogues such as Carfentanil, Acetyl Fentanyl, Butyryl Fentanyl, Remifentanil, Ocfentanil, Sufentanil, p-Fluoro Fentanyl, Furanyl Fentanyl, Valeryl Fentanyl, and 3-Methyl Fentanyl. The test strips are also able to be purchased separately.
Availability: Commercially Available
Iqbal Sunderani
iqbal@btnx.com
+1 905 944 9565 ext. 205
570 Hood Road, Unit 23
Markham, Ontario, L3R 4G7, Canada
Canada
|
|
www.btnx.com
| Manufacturer |
| BTNX Inc. |
| Product Name |
| Rapid Response Fentanyl (FYL) Forensic Test Kit |
| Detection Category |
| Chemical |
| Detection Principle |
| Colorimetric;Immunological; |
| Detection Method |
| Colorimetric;Lateral Flow Hand Held Immunoassay; |
| Application |
| Forensics;Screening; |
| Equipment Type |
| Instrument |
| Product Synopsis |
| The Rapid ResponseTM Fentanyl (FYL) Forensic Test Kit is a lateral flow chromatographic immunoassay for the qualitative detection of Fentanyl/Norfentanyl in liquid and powder substances at the cut-off concentration of 400 ng/mL. It is specific for Fentanyl screening with no significant cross reactivity to other opiates, such as Morphine and Heroin. The Rapid ResponseTM Fentanyl (FYL) Forensic Test Kit is able to detect Fentanyl and many other Fentanyl analogues such as Carfentanil, Acetyl Fentanyl, Butyryl Fentanyl, Remifentanil, Ocfentanil, Sufentanil, p-Fluoro Fentanyl, Furanyl Fentanyl, Valeryl Fentanyl, and 3-Methyl Fentanyl. The test strips are also able to be purchased separately. |
| Availability |
| Commercially Available |
| Technology Readiness Level (TRL) |
| 9 |
| Market Entry Date |
| 2008 |
| User Feedback Sources |
| MRIGlobal Experience |
| Supporting Documentation |
| Docs: BTNX - Rapid Response Fentanyl Test Kit.pdf; |
| Dimensions |
| 4.7 x 3.1 x 3.1 (12 x 8 x 8 cm) |
| Weight |
| <1 lb.; 3.5 oz (0.1 kg) |
| Power Requirements |
| None |
| Transportability |
| Pocket Size |
| Operating Conditions |
| 15 to 30°C (59 to 86°F) |
| Solvents/Reagents |
| Yes |
| Available Accessories |
| Swabs, test tubes, extraction buffers, workstation, product insert |
| Expected Operational Life |
| Single use item |
| Shelf Life |
| <5 years; 2 years |
| Unit Cost |
| $1 per test strip; $100 per test kit (100 strips) |
| Tested For Narcotics |
| Fentanyl, carfentanil, acetyl fentanyl, butyryl fentanyl, remifentanil, ocfentanil, sufentanil, p-fluoro fentanyl, furanyl fentanyl, valeryl fentanyl, 3-methyl fentanyl, norfentanyl, ketamine |
| Sample Introduction |
| Liquid;Solid; |
| Sensitivity/Detection Limits |
| %; ng; 2.5% solids, 400 ng liquids |
| False Positive Rates |
| 2% |
| False Negative Rates |
| 2% |
| Start Up Time (From Cold Start To Sample Ready) |
| Immediate |
| Response Time (Sample Application To Output) |
| <5 minutes |
| Total Run Time |
| 5 minutes |
| Other Operational Parameters |
| Interpretation of test results are: positive (one line), negative (two lines), invalid (no lines or no control line). |
| Training Required |
| <1 hour training |
| Training Available |
| No |
| Manuals Available |
| Instruction manual |
| Approved Vendors |
| DanceSafe, Test Kit Plus |