G908

DISCONTINUED; contact manufacturer for more information. G908 is a multi-column, 28-pound analyzer that combines the separation of gas chromatography, the compound selectivity of high pressure mass spectrometry (HPMS), and the analytical capabilities of FID and TCD, to provide qualitative and quantitative results.

Notify me on updates of this product 

Availability: Discontinued

John Kenneweg
Vice President of Sales, Safety & Security
john.kenneweg@908devices.com
+1 857 254 1500
27 Drydock Avenue
Boston, MA 02210
USA
Jim Kane
Director of Federal Programs
jim.kane@908devices.com
+1 757 746 5554
www.908devices.com
Manufacturer
908 Devices, Inc.
Product Name
G908
Detection Category
Chemical; Explosives;
Detection Principle
Spectrometry;Chromatography;
Detection Method
Ion Trap Mass Spectrometry;Gas Chromatography;
Application
Emergency Response;Forensics;Gas Analysis;Screening;
Product Model Number
G908
Equipment Type
Instrument
Product Synopsis
DISCONTINUED; contact manufacturer for more information. G908 is a multi-column, 28-pound analyzer that combines the separation of gas chromatography, the compound selectivity of high pressure mass spectrometry (HPMS), and the analytical capabilities of FID and TCD, to provide qualitative and quantitative results.
Availability
Discontinued
Technology Readiness Level (TRL)
9
Market Entry Date
Fall 2016
Dimensions
17 x 11 x 12 in (43 x 28 x 30 cm)
Weight
<30 lbs.; 28 lbs (12.7 kg)
Power Requirements
AC/DC Line Power; 24VDC from external power supply 100 to 240 VAC 50/60Hz
Noise Produced
Negligible
Transportability
Person-Portable
Ruggedness
Waterproof/Resistant; Dust; no shock or vibration testing
Operating Conditions
0 to 40°C (32 to 104°F)
Consumables
<$1,000/year; TD swabs
Solvents/Reagents
Carrier gas: helium, hydrogen or nitrogen
Data Analysis Support Equipment
Software
Communications Interface
USB
Expected Operational Life
Indefinite
Shelf Life
>5 years
Unit Cost
$50,000 to $65,000 depending on configuration
Tested For Chemical Agents
Yes
Tested For ITF-25 TIC/TIMs
Yes
Tested For Explosives
Yes
Tested For Narcotics
No
Other Chemical Targets
Yes
Adaptability
No
Sample Introduction
Liquid;Vapor/Gas;
Sensitivity/Detection Limits
ppb; ppm
Response Time (Sample Application To Output)
<5 minutes
Total Run Time
Minutes
Software Control
On-board, PC
Other Operational Parameters
Non-radioactive, internal glow discharge ionization (70eV), positive mode. Unit Mass Accuracy 1 m/z (40-250 m/z range with 100 ppm BTEX calibration standard). Unit mass range 15 to 450; Gas/Liquid injection.
Training Required
<1 day training
Training Available
Yes, 4 hours on-site
Manuals Available
Yes
Reachback Service
Yes
Warranties
Yes
ITAR/EAR Export Regulations
No
IATA Shipping Restrictions
Yes; Li-ion battery
Approved Vendors
Federal Resources
Please enter contact info.
Email Address: 
Company: 
Name: 
Title: