Flat Ferret

The FF-27 Tritium Contamination Monitors are portable gas flow proportional counters sensitive to betas of H-3. The windowless probes slide over flat surfaces on poly tape which gives repeatable spacing from surface as well as easy scratch-free movement. The poly tape is easily cleaned but may also be replaced if contaminated. P-10 counting gas utilized in the FF-27 probes is non-combustible and can be used in any environment.

Notify me on updates of this product 

Availability: Commercially Available

sales@overhoff.com
+1 513 248 2400
1160 US Route 50
Milford, Ohio 45150-9705
USA
www.overhoff.com
Manufacturer
Overhoff Technology Corporation
Product Name
Flat Ferret
Detection Category
Radiological/Nuclear;
Detection Principle
Geiger-Muller Detectors;
Detection Method
Geiger-Muller Detectors;
Application
Survey Meter;
Product Model Number
FF-27
Equipment Type
Instrument
Product Synopsis
The FF-27 Tritium Contamination Monitors are portable gas flow proportional counters sensitive to betas of H-3. The windowless probes slide over flat surfaces on poly tape which gives repeatable spacing from surface as well as easy scratch-free movement. The poly tape is easily cleaned but may also be replaced if contaminated. P-10 counting gas utilized in the FF-27 probes is non-combustible and can be used in any environment.
Availability
Commercially Available
Technology Readiness Level (TRL)
9
User Feedback Sources
Indifferent/No user feedback
Supporting Documentation
Docs: Flat Ferret.pdf;
Dimensions
10 x 4.1 x 7 in (25.4 x 10.4 x 17.8 cm)
Weight
<30 lbs.; 7 lbs (3.2 kg)
Power Requirements
Alkaline Battery; 2 9V batteries (300 hours)
Noise Produced
Negligible
Transportability
Handheld
Consumables
<$1,000/year; P-10 Gas container; lecture bottle 1 hour.
Solvents/Reagents
P-10 counting gas
Calibration Schedule
Internal
Suggested Routine Maintenance
Clean/replace poly tape
Available Accessories
Larger capacity gas container, clothing/equipment face plate
Communications Interface
None
Expected Operational Life
Indefinite
Shelf Life
>5 years
Unit Cost
$10,300.00
Radiological Targets
Tritium
Adaptability
No; cannot detect or add new targets
Sample Introduction
Radioactive Particle
Sensitivity/Detection Limits
Bq; 100 Bq/100 cm2,
Response Time (Sample Application To Output)
Continuous (<30 seconds)
Total Run Time
1 hour on lecture bottle
Alarm Capability
Audible and visual
Other Operational Parameters
35 cts/s above background
Training Required
<1 day training
Manuals Available
Yes
Please enter contact info.
Email Address: 
Company: 
Name: 
Title: