Gasmet DX Series
Gasmet™ DX4000 is a portable IR gas analyzer which is typically used in stack emissions monitoring, comparison measurements, catalytic process control and other applications where multiple gas compounds need to be accurately monitored in hot & humid sample gas. It is an ideal tool to measure trace concentrations of pollutants in wet, corrosive gas streams. Gasmet™ DX4015 is a portable IR gas analyzer for ambient air analysis. It is typically used in industrial hygiene and emergency situations. Up to 50 compounds can be measured simultaneously. Gasmet Technologies newest portable IR Gas Analyzer is the DX4040. The DX4040 comes with a 200 gas library for identification and accurate analysis of gases in part-per-million concentrations. Up to 25 gases can be measured simultaneously. Typical usage areas include industrial hygiene and emergency response situations.
Availability: Commercially Available
| Manufacturer |
| Gasmet Technologies, Inc. |
| Product Name |
| Gasmet DX Series |
| Detection Category |
| Chemical |
| Detection Principle |
| Molecular Spectroscopy; |
| Detection Method |
| Mid-infrared Spectroscopy; |
| Application |
| Gas Analysis; |
| Product Model Number |
| DX4000, DX4015, DX4040 |
| Equipment Type |
| Instrument |
| Product Synopsis |
| Gasmet™ DX4000 is a portable IR gas analyzer which is typically used in stack emissions monitoring, comparison measurements, catalytic process control and other applications where multiple gas compounds need to be accurately monitored in hot & humid sample gas. It is an ideal tool to measure trace concentrations of pollutants in wet, corrosive gas streams. Gasmet™ DX4015 is a portable IR gas analyzer for ambient air analysis. It is typically used in industrial hygiene and emergency situations. Up to 50 compounds can be measured simultaneously. Gasmet Technologies newest portable IR Gas Analyzer is the DX4040. The DX4040 comes with a 200 gas library for identification and accurate analysis of gases in part-per-million concentrations. Up to 25 gases can be measured simultaneously. Typical usage areas include industrial hygiene and emergency response situations. |
| Availability |
| Commercially Available |
| Technology Readiness Level (TRL) |
| 9 |
| User Feedback Sources |
| Hands-on experience |
| Dimensions |
| 15.3 x 17.5 x 6.3 in (39 x 45 x 16 cm) |
| Weight |
| <50 lbs.; 31 lbs (13.9 kg) |
| Power Requirements |
| Lithium Battery; AC/DC Line Power; Standard AC power: 100 to 115 or 230 V / 50 to 60 Hz, Lithium-ion battery (2 to 5 hours) |
| Noise Produced |
| Some noise generated by internal sample pump |
| Transportability |
| Person-Portable |
| Ruggedness |
| Waterproof/Resistant; Dust; IP 67 (DX4040). Sensitive to large fluctuations in temperature and high humidity |
| Operating Conditions |
| 0 to 40°C (32 to 150°F) Long term temperature: 5 to 30°C (41 to 86°F); 90%RH, non-condensing |
| Consumables |
| >$1,000/year; Nitrogen gas, 5.0 purity or higher |
| Calibration Schedule |
| Daily for zero calibration with nitrogen or air. |
| Available Accessories |
| Wheeled cart for the analyzer and laptop computer |
| Data Analysis Support Equipment |
| CALCMET software for Windows XP. |
| Communications Interface |
| Bluetooth, RS232 |
| Expected Operational Life |
| Indefinite |
| Shelf Life |
| >5 years |
| Unit Cost |
| $60,000.00 |
| Tested For Chemical Agents |
| VX, L, GB, GD, HD, GA, PS, DMMP, Chlorosoman, HCN, dimethyl phosphite, phosgene |
| Tested For ITF-25 TIC/TIMs |
| Few |
| Other Chemical Targets |
| H2O, CO2, CO, NO, NO2, N2O, SO2, NH3, CH4, HCl, HF, phosphine, sulfuryl fluoride, and different VOC´s. |
| Number of Library Entries |
| 50 to 200 |
| Adaptability |
| Yes; user can add new targets to detector with support |
| Sample Introduction |
| Vapor/Gas |
| Sensitivity/Detection Limits |
| ppm |
| Start Up Time (From Cold Start To Sample Ready) |
| 1 to 15 minutes |
| Response Time (Sample Application To Output) |
| <5 minutes; < 2 minutes |
| Total Run Time |
| < 2 minutes |
| Alarm Capability |
| Yes |
| Software Control |
| An external computer controls the GASMET through the operating software, “CALCMET” for Windows XP. |
| Other Operational Parameters |
| Simultaneous analysis of up to 50 gas compounds |
| Training Required |
| <1 week training |
| Training Available |
| Vendor training at customer or vendor facility |
| Manuals Available |
| Operator’s Manual, Software Manual |
| Reachback Service |
| No |
| Warranties |
| 1 year |
| ITAR/EAR Export Regulations |
| No |
| IATA Shipping Restrictions |
| Yes; Li-ion battery |
| Data File Type Format |
| SPE |