zNose Series
The Model 4300 zNose is a fully portable GC/SAW that can detect and analyze all types of vapors and identify traces of organic and chemical compounds quickly and accurately. It is designed with an internal helium cylinder (refillable), and can adapt to use disposable helium cylinders. The lithium ion battery is rechargeable, typically lasting 6 hours. The 4300 and 7100 models provide more quantitative results than the 4600, but require an external computer. The 7100 model is not portable.
Availability: Commercially available
Manufacturer |
| Electronic Sensor Technology, Inc. |
Product Name |
| zNose Series |
Detection Category |
| Chemical; Explosives |
Detection Principle |
| Chromatography; Sensors; |
Detection Method |
| Surface Acoustic Wave Detection; |
Application |
| Emergency Response; Gas Analysis; Screening; Bulk Analysis; |
Product Model Number |
| 4200, 4300, 4600, 7100 |
Equipment Type |
| Instrument |
Product Synopsis |
| The Model 4300 zNose is a fully portable GC/SAW that can detect and analyze all types of vapors and identify traces of organic and chemical compounds quickly and accurately. It is designed with an internal helium cylinder (refillable), and can adapt to use disposable helium cylinders. The lithium ion battery is rechargeable, typically lasting 6 hours. The 4300 and 7100 models provide more quantitative results than the 4600, but require an external computer. The 7100 model is not portable. |
Availability |
| Commercially available |
Technology Readiness Level (TRL) |
| 9 |
User Feedback Sources |
| Indifferent/No user feedback |
Dimensions |
| Head: 12.5 x 4.3 x 6.8 in (318 x 110 x 173 mm) Support: 12.5 x 9.7 x 5.8 in (318 x 247 x 148 mm) Charger: 13.5 x 9.7 x 3.7 in (343 x 247 x 94 mm) |
Weight |
| <50 lbs.; Head: 5.7 lbs (2.6 kg) Support: 18.7 lbs (8.5 kg) Charger: 7.7 lbs ( 3.5 kg) |
Power Requirements |
| Lithium Battery; AC/DC Line Power; Fuse: 4 amp, 250 V AC Battery Pack (4300 only): 28 V DC, 16 A Lithium-ion battery (6 hour use typical) |
Transportability |
| Person-Portable |
Ruggedness |
| None |
Operating Conditions |
| 0 to 40°C (32 to 105°F); 0 to 95%RH, non-condensing |
Consumables |
| >$1,000/year; Helium; Purge Dryer Refill; Purge Dryer Cartridge Kit; 1 µL syringe; Performance Evaluation Kit VOC (C6 – C14) |
Calibration Schedule |
| Yes |
Suggested Routine Maintenance |
| Occasional column replacement |
Available Accessories |
| Shipping case; Battery Assembly; DC/AC Inverter; Battery Charger |
Data Analysis Support Equipment |
| Laptop or PC |
Communications Interface |
| RS-232 between controller and 4300 Bluetooth |
Expected Operational Life |
| Indefinite |
Shelf Life |
| >5 years |
Unit Cost |
| $30,000.00 |
Maintenance Cost |
| Annual PM $4,300; Replacement column $995; Validation service: $1,495 |
Tested For Chemical Agents |
| GB and GD |
Tested For ITF-25 TIC/TIMs |
| Yes |
Tested For Explosives |
| TNT, TATP, C4, NG |
Other Chemical Targets |
| Volatile chemicals |
Adaptability |
| Yes; manufacturer can add new targets to detector |
Sample Introduction |
| Vapor/Gas |
Sensitivity/Detection Limits |
| ppb |
Known Interferents/Inhibitors |
| Gas chromatograph separates CAs from most interferents |
Start Up Time (From Cold Start To Sample Ready) |
| 15 minutes |
Response Time (Sample Application To Output) |
| <1 minute; 5 to 60 seconds |
Total Run Time |
| < 25 minutes |
Alarm Capability |
| Audible and visual |
Software Control |
| Product software needs an Intel Pentium 100 MHz or better processor with a minimum of 16 MB RAM, a 1GB Hard Drive, and any version of Microsoft Windows |
Training Required |
| <1 week training |
Training Available |
| Yes, classroom training for one student $2,400 |
Manuals Available |
| Yes |
Warranties |
| 1 year: parts only Hardware: 180 days for parts and labor Software: 180 days |
IATA Shipping Restrictions |
| Yes; Shipment of gas is by ground only |