Zr735
Zirconium Oxide Cell Sensor for measuring Oxygen in a high-purity inert gas application
Availability: Commercially Available
| Manufacturer |
| Servomex Group Limited |
| Product Name |
| Zr735 |
| Detection Category |
| Chemical |
| Detection Principle |
| Sensors; |
| Detection Method |
| Metal Oxide Semiconductor; |
| Application |
| Gas Analysis; |
| Product Model Number |
| Zr753 Trace |
| Equipment Type |
| Sensor |
| Product Synopsis |
| Zirconium Oxide Cell Sensor for measuring Oxygen in a high-purity inert gas application |
| Availability |
| Commercially Available |
| Technology Readiness Level (TRL) |
| 9 |
| User Feedback Sources |
| Indifferent/No user feedback |
| Dimensions |
| Control Electronics: 160 x 100 x 29mm |
| Weight |
| <1 lb.; Sensor Only: 270gms (9.5 oz.) including two gas connectors; 13gms for Control Electronics |
| Power Requirements |
| AC/DC Line Power; 15V-0-15V ac, 47Hz-65Hz, min 12V-0-12V, max
21V-0-21V, Transformer winding rated at 2.7A rms |
| Transportability |
| Pocket Size |
| Ruggedness |
| Vibration/Shock; Meets requirements of BS EN 60068-2-6:1996 (IEC 68-2-6), BS EN 60068-2-27:1993 (IEC 68-2-27), IEC 68-2-34, BS EN 150 80601-2-12_2011, ISO 80601-2-58 |
| Operating Conditions |
| RH 0 to 90% |
| Communications Interface |
| Analog, The sensor output is amplified by the
Control Electronics to give an output from
the kit equal to: (2.3E + 0.106)V
The output from the sensor
follows the Nernst equation:
Where E = sensor output (V), R = Gas
Constant, T = Electrode temperature (K),
F = Faraday Constant
PO2’’ = Oxygen partial pressure of reference
gas, PO2’ = Oxygen partial pressure of
sample gas |
| Other Chemical Targets |
| O2 |
| Sample Introduction |
| Vapor/Gas |
| Sensitivity/Detection Limits |
| ppm |
| Start Up Time (From Cold Start To Sample Ready) |
| 3 to 5 minutes |
| Response Time (Sample Application To Output) |
| Continuous (<30 seconds); <8 secs for a step change from 2ppm(v) to 10ppm(v) O2 |
| Training Required |
| <1 day training |