Detective-EX-100T/Detective-DX-100T
The Ortec Detective is a battery operated PINS system with neutron source for identification of chemicals within thick steel containers or other suspect packaging. Two models are available. The Detective-EX-100T has a large 65 mm x 50 mm HPGe detector for gamma and a neutron detector. Detective-DX-100T has a large 65 mm x 50 mm HPGe detector for gamma detection only.
Availability: DISCONTINUED
Manufacturer |
| Ortec (Ametek) |
Product Name |
| Detective-EX-100T/Detective-DX-100T |
Detection Category |
| Radiological/Nuclear |
Detection Principle |
| Semiconductors; |
Detection Method |
| Semiconductors; |
Application |
| Survey Meter; Identifiers; |
Product Model Number |
| DETEX-100T-GP |
Equipment Type |
| Instrument |
Product Synopsis |
| The Ortec Detective is a battery operated PINS system with neutron source for identification of chemicals within thick steel containers or other suspect packaging. Two models are available. The Detective-EX-100T has a large 65 mm x 50 mm HPGe detector for gamma and a neutron detector. Detective-DX-100T has a large 65 mm x 50 mm HPGe detector for gamma detection only. |
Availability |
| DISCONTINUED |
Technology Readiness Level (TRL) |
| 9 |
User Feedback Sources |
| Indifferent/No user feedback |
Dimensions |
| 15.5 x 7.2 x 13.75 in (39.4 x 18.3 x 34.9 cm) |
Weight |
| <30 lbs.; 26.3 lbs (12 kg) |
Power Requirements |
| AC/DC Line Power; Lithium Battery; Docking station or attached to the compact AC/Power Adapter/Charger (PAC); Battery (3 hours) |
Transportability |
| Handheld |
Operating Conditions |
| -10 to 40°C (14 to 105°F), <90%RH, non-condensing |
Consumables |
| None |
Calibration Schedule |
| Auto-calibration with K-40 naturally occuring peak |
Available Accessories |
| Check Source, transport case, external battery pack |
Data Analysis Support Equipment |
| Onboard processing, with sync possible to external computer |
Communications Interface |
| USB/SD Card Slot, wireless, GPS |
Expected Operational Life |
| Indefinite |
Shelf Life |
| >5 years |
Unit Cost |
| $100,000.00 |
Radiological Targets |
| Gamma, neutrons |
Number of Library Entries |
| 150 |
Adaptability |
| No; cannot detect or add new targets |
Sample Introduction |
| Radioactive Particle |
Sensitivity/Detection Limits |
| µSv; <0.05 µSv/h |
Start Up Time (From Cold Start To Sample Ready) |
| 12 hours |
Response Time (Sample Application To Output) |
| Continuous (<30 seconds); 8 seconds |
Total Run Time |
| Continuous |
Alarm Capability |
| Yes |
Software Control |
| PC, remote |
Other Operational Parameters |
| GPS; ANSI N42-34 |
Training Required |
| <1 day training |
Training Available |
| Yes |
Manuals Available |
| Yes |