Applied Biosystems 7500 Fast Real-Time PCR System

The Applied Biosystems 7500 Fast Real-Time PCR System is a real-time PCR system which enables high-speed thermal cycling with run times for quantitative real-time PCR applications of fewer than 40 minutes. The system is an open platform and can be used to analyze samples using custom made or COTs-based PCR assays. The 7500 Fast system is batch-cable and can analyzed a batch of 96 samples per run. It is used for gene expression analysis, pathogen quantitation, SNP genotyping, isothermal and +/– assays utilizing internal positive controls.

Notify me on updates of this product 

Availability: Commercially Available

Dan Davis
Product Manager
daniel.davis2@thermofisher.com
978-513-9717
2 Radcliff Road
Tewksbury, MA 01876
USA
www.thermoscientific.com
Manufacturer
Thermo Fisher Scientific, Inc.
Product Name
Applied Biosystems 7500 Fast Real-Time PCR System
Detection Category
Biological;
Detection Principle
Molecular;
Detection Method
Polymerase Chain Reaction;
Application
Laboratory Analysis;
Product Model Number
4351106
Equipment Type
Instrument
Product Synopsis
The Applied Biosystems 7500 Fast Real-Time PCR System is a real-time PCR system which enables high-speed thermal cycling with run times for quantitative real-time PCR applications of fewer than 40 minutes. The system is an open platform and can be used to analyze samples using custom made or COTs-based PCR assays. The 7500 Fast system is batch-cable and can analyzed a batch of 96 samples per run. It is used for gene expression analysis, pathogen quantitation, SNP genotyping, isothermal and +/– assays utilizing internal positive controls.
Availability
Commercially Available
Technology Readiness Level (TRL)
9
User Feedback Sources
Indifferent/No user feedback
Dimensions
34 cm (13.39 in) (W) x 49 cm (19.29 in) (H) x 45 cm (17.72 in) (D)
Weight
>50 Lbs.; 34 kg (75 Lbs.)
Power Requirements
AC/DC Line Power; 120V, 50/60Hz, 8A, 950W
Transportability
Benchtop/Fixed
Ruggedness
Indoor use only
Operating Conditions
Temperature: 15 to 30 °C (59 to 86 °F). Maximum change of less than 15 degrees Celsius (27 °F) per 24 hrs. Humidity: 20 to 80% relative humidity, noncondensing.
Consumables
>$1,000/year; 96 well plates
Calibration Schedule
Monthly/6 months or if moved
Suggested Routine Maintenance
Weekly, monthly and 6 month maintenance
Data Analysis Support Equipment
Computer
Communications Interface
USB
Unit Cost
$35,000.00
Maintenance Cost
$4,800 annual service contract
Tested For Biological Agents
B.anthracis, F. tularensis, Y. pestis
Other Biological Targets
VEEV, Escherichia coli O157:H7, SARS-CoV-2 (COVID-19), etc. Almost any custom or COTS assays can be analyzed with this instrument
Adaptability
Yes; user can add new targets to detector with support
Sample Introduction
Liquid
Sensitivity/Detection Limits
copies/mL; Down to 1 copy of human RNAse P gene
Start Up Time (From Cold Start To Sample Ready)
< 1 minute
Response Time (Sample Application To Output)
>15 minutes; 36 minutes (Fast), 2 hours (Standard)
Total Run Time
36 minutes (Fast), 2 hours (Standard)
Alarm Capability
Visual
Software Control
PC
Training Required
>1 week training
Training Available
Yes; in-person, webinars, videos
Manuals Available
Yes
Warranties
1 year
Data File Type Format
SDS