MINICAMS 3500/2010
The MINICAMS® Series is an automatic, near-real-time continuous air monitoring system using gas chromatography and sample collection with a solid-adsorbent preconcentrator or fixed-volume sample loop. The MINICAMS collects an air sample, performs an analysis, and reports the result. Three different Plug-In GC Modules allow configurations to maximize selectivity and sensitivity for monitoring all chemical warfare (CW) agents. Multiple MINICAMS can be networked to provide a facility-wide Air Monitoring System.
Availability: DISCONTINUED
Greg Houston
Sales Manager
+1 205 733 6900 Ext. 204
2148 Pelham Parkway, Building 400
Pelham, AL 35124
USA
|
|
www.oico.com
| Manufacturer |
| OI Analytical |
| Product Name |
| MINICAMS 3500/2010 |
| Detection Category |
| Chemical |
| Detection Principle |
| Chromatography; Sensors; |
| Detection Method |
| Flame Photometric Detection (FPD); Pulsed Flame Photometric Detector (PFPD); Flame Ionization Detection (FID); Photoionization Detection; Gas Chromatography; |
| Application |
| Gas Analysis; Laboratory Analysis; |
| Equipment Type |
| Instrument |
| Product Synopsis |
| The MINICAMS® Series is an automatic, near-real-time continuous air monitoring system using gas chromatography and sample collection with a solid-adsorbent preconcentrator or fixed-volume sample loop. The MINICAMS collects an air sample, performs an analysis, and reports the result. Three different Plug-In GC Modules allow configurations to maximize selectivity and sensitivity for monitoring all chemical warfare (CW) agents. Multiple MINICAMS can be networked to provide a facility-wide Air Monitoring System. |
| Availability |
| DISCONTINUED |
| Technology Readiness Level (TRL) |
| 9 |
| User Feedback Sources |
| Indifferent/No user feedback |
| Dimensions |
| 12 x 12 x 10 in (305 x 305 x 254 mm) |
| Weight |
| <30 lbs.; 18 lbs (8.2 kg) |
| Power Requirements |
| AC/DC Line Power; 120 VAC |
| Transportability |
| Benchtop/Fixed; Small Footprint |
| Ruggedness |
| None; Not militarized, drop sensitive |
| Operating Conditions |
| 5 to 40°C (41 to 104°F) |
| Consumables |
| >$1,000/year; Air, high purity hydrogen for FPD and PFPD, helium carrier gas |
| Solvents/Reagents |
| None |
| Calibration Schedule |
| Calibrate for specific needs |
| Suggested Routine Maintenance |
| Columns, detectors, concentrators |
| Data Analysis Support Equipment |
| Continuous sampling system Stream selection systems Heated sample lines Sample flow controller Low volume sampler |
| Communications Interface |
| RS232 RS485 |
| Shelf Life |
| >5 years |
| Unit Cost |
| $44,820.00 |
| Maintenance Cost |
| $2,500 per year |
| Tested For Chemical Agents |
| Nerve agent, blister agents |
| Tested For ITF-25 TIC/TIMs |
| Phosgene, cyanogen chloride, chloropicrin, methylene chloride, vinyl chloride, phosphine, DMMP, malathion benzene, nitrobenzene, toluene, ethylene oxide, acetonitrile |
| Other Chemical Targets |
| Yes |
| Adaptability |
| Yes |
| Sample Introduction |
| Vapor/Gas |
| Sensitivity/Detection Limits |
| ng |
| Start Up Time (From Cold Start To Sample Ready) |
| 10 minutes |
| Response Time (Sample Application To Output) |
| Continuous (<30 seconds) |
| Alarm Capability |
| Integral audible and visual alarm indicators |
| Software Control |
| Chrom-NET |
| Other Operational Parameters |
| Programmable for specific needs |
| Training Required |
| <1 week training |
| Training Available |
| Different training options available |
| Warranties |
| 1 year parts and labor |
| IATA Shipping Restrictions |
| Yes; Compressed gas cylinders must be shipped by ground only per IATA regulations. |