ppbRAE 3000+

The ppbRAE 3000+ is a handheld multigas chemical detection and identification system . The ppbRAE 3000 is a photoionization detector (PID) with a range of 1 ppb to 10,000 ppm. The unit is fully portable and meant for field use. The unit can be equipped with a 10.6 eV or 9.8 eV lamp.

Notify me on updates of this product 

Availability: Commercially Available

Tom Kochevar
detectgas@honeywell.com
+1 877 723 2878
3775 North First Street
San Jose, CA 65134
USA
www.honeywellanalytics.com
Manufacturer
Honeywell Analytics
Product Name
ppbRAE 3000+
Detection Category
Chemical
Detection Principle
Sensors;
Detection Method
Photoionization Detection;
Application
Emergency Response; Gas Analysis;
Equipment Type
Instrument
Product Synopsis
The ppbRAE 3000+ is a handheld multigas chemical detection and identification system . The ppbRAE 3000 is a photoionization detector (PID) with a range of 1 ppb to 10,000 ppm. The unit is fully portable and meant for field use. The unit can be equipped with a 10.6 eV or 9.8 eV lamp.
Availability
Commercially Available
Technology Readiness Level (TRL)
9
User Feedback Sources
MRIGlobal hands-on experience; positive user feedback
Dimensions
10 x 3 x 2.5 in (25 x 8 x 6.4 cm)
Weight
<5 Lbs.; 26 oz. (738 g)
Power Requirements
Lithium Battery; Alkaline Battery; Rechargeable, external field-replaceable Lithium-ion battery pack, Alkaline battery adapter
Transportability
Pocket Size
Ruggedness
Extreme Temperatures/Humidity; Waterproof/Resistant; Dust; IP67, protective rubber boot
Operating Conditions
–20 to 50°C (–4 to 122°F); 0 to 95% RH, non-condensing
Consumables
<$1,000/year; Porous metal filters and O-rings, 100 ppm isobutylene calibration gas, 10.6 eV lamp
Solvents/Reagents
Isobutylene calibration gas
Calibration Schedule
Two-point or three-point calibration for zero and span
Suggested Routine Maintenance
O-ring replacement, lamp replacement
Available Accessories
Hard transport case, charging/download cradle, Organic vapor zeroing kit , Gas outlet port adapter and tubing
Communications Interface
Bluetooth Built-in RF modem
Expected Operational Life
Indefinite
Shelf Life
>5 years
Unit Cost
$5,000.00
Maintenance Cost
Annual maintenance service, cost unspecified
Other Chemical Targets
Volatile organic compounds
Adaptability
No; cannot detect or add new targets
Sample Introduction
Vapor/Gas
Sensitivity/Detection Limits
ppm; ppb
Start Up Time (From Cold Start To Sample Ready)
< 5 minutes
Response Time (Sample Application To Output)
Continuous (<30 seconds); 10 seconds
Total Run Time
< 5 minutes
Alarm Capability
Audible and visual
Software Control
On-board
Training Required
<1 day training
Training Available
Yes
Manuals Available
Operator’s Manual, Maintenance Manual
Warranties
4 years repair and replacement guarantee, 3 years on 10.6 eV lamp
IATA Shipping Restrictions
Yes; Lithium-ion battery restrictions and calibration gas IATA regulations
Please enter contact info.
Email Address: 
Company: 
Name: 