ppbRAE 3000+
The ppbRAE 3000+ is a handheld multigas chemical detection and identification system . The ppbRAE 3000 is a photoionization detector (PID) with a range of 1 ppb to 10,000 ppm. The unit is fully portable and meant for field use. The unit can be equipped with a 10.6 eV or 9.8 eV lamp.
Availability: Commercially Available
| Manufacturer |
| Honeywell Analytics |
| Product Name |
| ppbRAE 3000+ |
| Detection Category |
| Chemical; |
| Detection Principle |
| Sensors; |
| Detection Method |
| Photoionization Detection; |
| Application |
| Emergency Response; Gas Analysis; |
| Equipment Type |
| Instrument |
| Product Synopsis |
| The ppbRAE 3000+ is a handheld multigas chemical detection and identification system . The ppbRAE 3000 is a photoionization detector (PID) with a range of 1 ppb to 10,000 ppm. The unit is fully portable and meant for field use. The unit can be equipped with a 10.6 eV or 9.8 eV lamp. |
| Availability |
| Commercially Available |
| Technology Readiness Level (TRL) |
| 9 |
| User Feedback Sources |
| MRIGlobal hands-on experience; positive user feedback |
| Supporting Documentation |
| Docs: Datasheet_ppbRAE2030002B_ES.pdf; |
| Dimensions |
| 10 x 3 x 2.5 in (25 x 8 x 6.4 cm) |
| Weight |
| <5 Lbs.; 26 oz. (738 g) |
| Power Requirements |
| Lithium Battery; Alkaline Battery; Rechargeable, external field-replaceable Lithium-ion battery pack, Alkaline battery adapter |
| Transportability |
| Pocket Size |
| Ruggedness |
| Extreme Temperatures/Humidity; Waterproof/Resistant; Dust; IP67, protective rubber boot |
| Operating Conditions |
| –20 to 50°C (–4 to 122°F); 0 to 95% RH, non-condensing |
| Consumables |
| <$1,000/year; Porous metal filters and O-rings, 100 ppm isobutylene calibration gas, 10.6 eV lamp |
| Solvents/Reagents |
| Isobutylene calibration gas |
| Calibration Schedule |
| Two-point or three-point calibration for zero and span |
| Suggested Routine Maintenance |
| O-ring replacement, lamp replacement |
| Available Accessories |
| Hard transport case, charging/download cradle, Organic vapor zeroing kit , Gas outlet port adapter and tubing |
| Communications Interface |
| Bluetooth Built-in RF modem |
| Expected Operational Life |
| Indefinite |
| Shelf Life |
| >5 years |
| Unit Cost |
| $5,000.00 |
| Maintenance Cost |
| Annual maintenance service, cost unspecified |
| Other Chemical Targets |
| Volatile organic compounds |
| Adaptability |
| No; cannot detect or add new targets |
| Sample Introduction |
| Vapor/Gas |
| Sensitivity/Detection Limits |
| ppm; ppb |
| Start Up Time (From Cold Start To Sample Ready) |
| < 5 minutes |
| Response Time (Sample Application To Output) |
| Continuous (<30 seconds); 10 seconds |
| Total Run Time |
| < 5 minutes |
| Alarm Capability |
| Audible and visual |
| Software Control |
| On-board |
| Training Required |
| <1 day training |
| Training Available |
| Yes |
| Manuals Available |
| Operator’s Manual, Maintenance Manual |
| Warranties |
| 4 years repair and replacement guarantee, 3 years on 10.6 eV lamp |
| IATA Shipping Restrictions |
| Yes; Lithium-ion battery restrictions and calibration gas IATA regulations |