DL-102+
The DL-102+ Snap-On PID™ Photoionization Analyzer is a single piece, hand-held unit for the measurement of organic and inorganic gas phase species. The Model 102+ is a PID based VOC analyzer that can accommodate 0 to 3 additional sensors (over 30 sensors to choose from). Includes operator's manual and rechargeable nickel metal hydride battery.
Availability: Commercially Available
Manufacturer |
| PID Analyzers LLC |
Product Name |
| DL-102+ |
Detection Category |
| Chemical |
Detection Principle |
| Molecular Spectroscopy; Sensors; |
Detection Method |
| Photoionization Detection; Electrochemical Detection; Infrared (IR) Spectroscopy ; |
Application |
| Emergency Response; Gas Analysis; |
Equipment Type |
| Instrument |
Product Synopsis |
| The DL-102+ Snap-On PID™ Photoionization Analyzer is a single piece, hand-held unit for the measurement of organic and inorganic gas phase species. The Model 102+ is a PID based VOC analyzer that can accommodate 0 to 3 additional sensors (over 30 sensors to choose from). Includes operator's manual and rechargeable nickel metal hydride battery. |
Availability |
| Commercially Available |
Technology Readiness Level (TRL) |
| 9 |
User Feedback Sources |
| Indifferent/No user feedback |
Dimensions |
| 8 x 3 x 2.25 in (203 x 76 x 57 mm) |
Weight |
| <5 lbs.; 1.9 lbs (0.9 kg) |
Power Requirements |
| AC/DC Line Power; 12 V DC charger with cigarette lighter adapter |
Transportability |
| Handheld |
Operating Conditions |
| 5 to 40°C (41 to 104°F); 0 to 95% RH non-condensing |
Consumables |
| <$1,000/year; Calibration gas, calibration kit: $250 |
Calibration Schedule |
| Automatic electronic zero provides a reliable calibration point so that when calibrated with a span gas, this is a two-point calibration. Calibration only takes 10 seconds. Optional calibration kit can be purchased |
Suggested Routine Maintenance |
| Clean and calibrate service/annually: $229 A calibration kit can be purchased for $250. |
Data Analysis Support Equipment |
| Data logging software |
Communications Interface |
| RS 232 |
Expected Operational Life |
| Indefinite |
Shelf Life |
| >5 years |
Unit Cost |
| $3,850.00 |
Maintenance Cost |
| $495 lamp replacement (every 2 to 3 years). Sensors can be added—prices from $175 to $750. |
Tested For ITF-25 TIC/TIMs |
| VOCs |
Number of Library Entries |
| Not Specified |
Sample Introduction |
| Vapor/Gas |
Sensitivity/Detection Limits |
| ppm |
Known Interferents/Inhibitors |
| Solvents may cause interference |
Start Up Time (From Cold Start To Sample Ready) |
| 10 to 15 seconds |
Response Time (Sample Application To Output) |
| <1 minute; 1 to 60 seconds |
Total Run Time |
| < 1 minute |
Alarm Capability |
| Audible |
Software Control |
| Downloading software is included with model for Windows HyperTerminal |
Training Required |
| <1 day training |
Training Available |
| No |
Manuals Available |
| Operator’s manual |
Warranties |
| 1 year |
IATA Shipping Restrictions |
| Yes; Calibration gases fall under IATA shipping regulations |