InPhochelle

A Raman instrument, which uses one of two laser options a 670 nm excitation and a 785 nm excitation, to measure a sample. A probe is either placed in the sample or up to the sample.

Notify me on updates of this product 

Availability: Commercially Available

Rob Forney
rob.forney@inphotonics.com
+1 781 440 0202
111 Downey Street
Norwood, MA 02062
USA
www.inphotonics.com
Manufacturer
InPhotonics, Inc.
Product Name
InPhochelle
Detection Category
Chemical; Explosives
Detection Principle
Molecular Spectroscopy;
Detection Method
Raman Spectroscopy;
Application
Emergency Response; Screening; Bulk Analysis;
Product Model Number
InPhochelle 670, InPhochelle 785
Equipment Type
Instrument
Product Synopsis
A Raman instrument, which uses one of two laser options a 670 nm excitation and a 785 nm excitation, to measure a sample. A probe is either placed in the sample or up to the sample.
Availability
Commercially Available
Technology Readiness Level (TRL)
9
Market Entry Date
2014
User Feedback Sources
Indifferent/No user feedback
Dimensions
21 x 10.5 x 5 in (53 x 27 x 13 cm)
Weight
<30 lbs.; 17 lbs (7.7 kg)
Power Requirements
AC/DC Line Power; AC Power
Transportability
Person-Portable
Ruggedness
None
Operating Conditions
0 to 45°C (32 to 113°F)
Consumables
None
Calibration Schedule
Fixed optics provide permanent calibration under normal operating conditions. Intensity correction performed automatically through software.
Suggested Routine Maintenance
User can perform basic cleaning; Replace laser every 4 to 5 years
Available Accessories
Extra libraries available ($1000 each), RamanProbeTM, Class I sample holder
Data Analysis Support Equipment
Basic system includes data acquisition software for control of spectrometer. GRAMS AI manipulation software, PC data station running Windows XP or Windows 7 required for operation.
Communications Interface
USB
Expected Operational Life
Indefinite
Shelf Life
>5 years
Unit Cost
$45,000.00
Maintenance Cost
Extra libraries $1000 each
Tested For Chemical Agents
Yes
Tested For ITF-25 TIC/TIMs
Yes
Tested For Explosives
Yes
Tested For Narcotics
Yes
Other Chemical Targets
Yes
Number of Library Entries
243
Adaptability
Yes; user can add new targets in the field
Sample Introduction
Liquid; Solid
Sensitivity/Detection Limits
Bulk; mg; g; %
Known Interferents/Inhibitors
Analysis of mixtures can produce erroneous search results unless the library contains a reference spectrum of the mixture itself. However, it is still capable of detecting a target in the presence of an unknown mixture. Fluorescence, either from the unknown, or a component or interferent in the unknown, can produce intense interference. The technique is sensitive to ambient infrared radiation.
Start Up Time (From Cold Start To Sample Ready)
5 minutes
Response Time (Sample Application To Output)
Continuous (<30 seconds); 1 to 10 seconds
Total Run Time
5 minutes
Software Control
On-board, PC
Other Operational Parameters
Spectral range: 100 to 3500 cm-1 measured in a single exposure; TE cooled detector (-55°C); Excitation wavelength: 670 nm or 785 nm
Training Required
<1 day training
Training Available
Yes
Manuals Available
Yes
Warranties
1 year parts and labor
IATA Shipping Restrictions
No