Gasmet DX4000
Gasmet™ DX4000 is a portable IR gas analyzer which is typically used in stack emissions monitoring, comparison measurements, catalytic process control and other applications where multiple gas compounds need to be accurately monitored in hot & humid sample gas. It is an ideal tool to measure trace concentrations of pollutants in wet, corrosive gas streams. Sample cell absorption path length is selected according to the application.
Availability: Commercially Available
| Manufacturer |
| Gasmet Technologies, Inc. |
| Product Name |
| Gasmet DX4000 |
| Detection Category |
| Chemical |
| Detection Principle |
| Molecular Spectroscopy; |
| Detection Method |
| Mid-infrared Spectroscopy; |
| Application |
| Gas Analysis; |
| Product Model Number |
| DX4000 |
| Equipment Type |
| Instrument |
| Product Synopsis |
| Gasmet™ DX4000 is a portable IR gas analyzer which is typically used in stack emissions monitoring, comparison measurements, catalytic process control and other applications where multiple gas compounds need to be accurately monitored in hot & humid sample gas. It is an ideal tool to measure trace concentrations of pollutants in wet, corrosive gas streams. Sample cell absorption path length is selected according to the application. |
| Availability |
| Commercially Available |
| Technology Readiness Level (TRL) |
| 9 |
| User Feedback Sources |
| Indifferent/No user feedback |
| Dimensions |
| 15.3 x 17.5 x 6.3 in (389 x 445 x 160 mm) |
| Weight |
| <50 Lbs.; 31 Lbs. (13.9 kg) |
| Power Requirements |
| AC/DC Line Power; Standard AC power: 100-115 or 230 V / 50 -60 Hz |
| Transportability |
| Person-Portable |
| Ruggedness |
| None |
| Operating Conditions |
| Short term temperature: 0 to 40°C (32 to 150°F) Long term temperature: 5 to 30°C (41 to 86°F); 90%RH, non-condensing |
| Consumables |
| >$1,000/year; Nitrogen gas, 5.0 purity or higher |
| Solvents/Reagents |
| None |
| Calibration Schedule |
| Zero point calibration performed using nitrogen (4.0 or higher N2 recommended) immediately before sampling and no more than once every 24 hours |
| Available Accessories |
| Can be coupled with the portable sampling system. |
| Data Analysis Support Equipment |
| Sample Cell Portable Sampling System Calibrator Pressure measurement: Inside sample cell Analog Signals (ext PC): ADAM 5000/TCP Sample cell gaskets: Teflon® coated Viton® or Kalrez® -Trolley: wheeled cart for the analyzer and laptop computer |
| Communications Interface |
| Bluetooth, RS232 Analyzer is connected to an external computer via RS- 232C cable. The external computer controls the instrument. The system operating software is CALCMET for Windows XP. |
| Shelf Life |
| >5 years |
| Unit Cost |
| $60,000.00 |
| Tested For ITF-25 TIC/TIMs |
| Few |
| Other Chemical Targets |
| H2O, CO2, CO, NO, NO2, N2O, SO2, NH3, CH4, HCl, HF and different VOC´s. |
| Adaptability |
| Yes; user can add new targets to detector with support |
| Sample Introduction |
| Vapor/Gas |
| Sensitivity/Detection Limits |
| ppb; ppm |
| Start Up Time (From Cold Start To Sample Ready) |
| 1 to15 minutes |
| Response Time (Sample Application To Output) |
| <5 minutes; < 2 minutes |
| Total Run Time |
| < 2 minutes |
| Alarm Capability |
| Yes |
| Software Control |
| PC; An external computer controls the GASMET through the operating software, “CALCMET” for Windows XP. |
| Other Operational Parameters |
| Simultaneous analysis of up to 50 gas compounds |
| Training Required |
| <1 day training |
| Training Available |
| Vendor training at customer or vendor facility |
| Manuals Available |
| Operator’s Manual Software Manual |
| Warranties |
| 1 year |