Canary-Two
Defiant’s Canary-Two™ combines three micro-fabricated components: a chemical pre-concentrator, a gas chromatographic column and a SAW micro-balance. A chemical coating on the preconcentrator ignores potentially interfering chemicals while the GC column separates target chemicals for unique identification. The system identifies agents in air and reports their concentration.
Availability: DISCONTINUED
Douglas Adkins
(505) 553-4418
3825 Academy Parkway South NE
Albuquerque, NM 87109-4410
|
|
www.defiant-tech.com
Manufacturer |
| Defiant Technologies |
Product Name |
| Canary-Two |
Detection Category |
| Chemical |
Detection Principle |
| Chromatography; Sensors; |
Detection Method |
| Gas Chromatography;Surface Acoustic Wave Detection; |
Application |
| Gas Analysis; Emergency Response; |
Equipment Type |
| Instrument |
Product Synopsis |
| Defiant’s Canary-Two™ combines three micro-fabricated components: a chemical pre-concentrator, a gas chromatographic column and a SAW micro-balance. A chemical coating on the preconcentrator ignores potentially interfering chemicals while the GC column separates target chemicals for unique identification. The system identifies agents in air and reports their concentration. |
Availability |
| DISCONTINUED |
Technology Readiness Level (TRL) |
| 9 |
User Feedback Sources |
| Indifferent/No user feedback |
Dimensions |
| 7.5 x 5.0625 x 2.125 in (19.05 x 12.86 x 5.398 cm) |
Weight |
| <5 lbs.; 26 oz (.737 kg) |
Power Requirements |
| Lithium Battery; AC/DC Line Power; Lithium polymer (LiPo) battery life in continuous use, A/C adapter and battery recharger. |
Transportability |
| Handheld |
Ruggedness |
| None |
Calibration Schedule |
| Should be calibrated every few months for optimum performance |
Suggested Routine Maintenance |
| Pump replacement when needed. |
Communications Interface |
| A serial interface cable is provided, a USB adapter cable is optional. |
Expected Operational Life |
| 4 hour battery life |
Unit Cost |
| $28,000.00 |
Maintenance Cost |
| Vendor offers maintenance contract for ¼ sales cost (covering calibration, firmware upgrades, cleaning, and replacing damaged parts). |
Tested For Chemical Agents |
| Sarin, Soman, Tabun, Cyclosarin, and Sulfur mustard. |
Tested For ITF-25 TIC/TIMs |
| Benzene, ethyl benzene, TCE, PCE, xylenes (o and p/m), toluene, and vinyl chloride. |
Sample Introduction |
| Vapor/Gas |
Sensitivity/Detection Limits |
| ppb |
Known Interferents/Inhibitors |
| Vendor reports that a chemical coating on the preconcentrator ignores interfering chemicals while collecting target analytes. Selectivity for phosphonates >6000 to 1 over diesel and >500,000 to 1 over humidity. |
Start Up Time (From Cold Start To Sample Ready) |
| 120 seconds |
Response Time (Sample Application To Output) |
| <5 minutes; 140 seconds |
Alarm Capability |
| Yes |
Software Control |
| Serial output chromatographic data. |
Training Required |
| <1 day training |
Training Available |
| On request |
Manuals Available |
| Yes |
Warranties |
| Vendor offers maintenance contract for ¼ sales cost (covering calibration, firmware upgrades, cleaning, and replacing damaged parts). |
IATA Shipping Restrictions |
| Yes; Lithium ion battery |