CT-1128

The CT-1128 is a lightweight, ruggedized, field serviceable GC/MS designed for environmental assessments, arms control and nonproliferation activities, and detection and identification of hazardous substances. The MS is a repackaged Agilent 5973N with two turbopumps for faster pump down.

Notify me on updates of this product 

Availability: DISCONTINUED

Neal Werner
+1 727 547 0600
7887 Bryan Dairy Road, Suite 100
Largo, Florida 33777-1452
www.contech.com
Manufacturer
Constellation Technology Corporation
Product Name
CT-1128
Detection Category
Chemical
Detection Principle
Spectrometry;
Detection Method
Quadrupole Mass Spectrometry;
Application
Gas Analysis; Laboratory Analysis;
Equipment Type
Instrument
Product Synopsis
The CT-1128 is a lightweight, ruggedized, field serviceable GC/MS designed for environmental assessments, arms control and nonproliferation activities, and detection and identification of hazardous substances. The MS is a repackaged Agilent 5973N with two turbopumps for faster pump down.
Availability
DISCONTINUED
Technology Readiness Level (TRL)
9
User Feedback Sources
Indifferent/No user feedback
T&E Reports and Standards
Reports: CT1128_Final_128.pdf; CT1128_Final_128.pdf; CT1128_Final_128.pdf;
Dimensions
15 x 23 x 15 in (38 x 28.4 x 38 cm)
Weight
>50 lbs.; 75 lbs. (34 kg)
Power Requirements
AC/DC Line Power; 120 V AC, 5 A peak power
Transportability
Mounted
Ruggedness
Vibration/Shock; Shock-mounted for use in vehicle.
Operating Conditions
32 to 110°C
Consumables
>$1,000/year; Typical GC supplies (ferrules, liners, etc.). Uses hydrogen carrier gas supply from internal canister; 160 hr operating time per canister.
Calibration Schedule
Yes
Suggested Routine Maintenance
Typical for GC/MS
Communications Interface
Laptop
Unit Cost
$139,500.00
Maintenance Cost
$5 to $10K per year
Tested For Chemical Agents
Yes. Tested at SBCCOM.
Tested For ITF-25 TIC/TIMs
Yes
Tested For Explosives
Yes
Other Chemical Targets
Yes
Sample Introduction
Liquid; Solid
Sensitivity/Detection Limits
%; mg
Known Interferents/Inhibitors
GC/MS is highly selective for volatile and semivolatile organic compounds when used by a skilled operator.
Start Up Time (From Cold Start To Sample Ready)
60 to 90 minutes from pumpdown to operational ready for analysis.
Response Time (Sample Application To Output)
>15 minutes; 5 to 30 minutes (GC runtime)
Alarm Capability
Audible and visual alarms
Software Control
Laptop computer with Agilent Chemstation software. NIST/AMSID compatible.
Other Operational Parameters
Same as Agilent 5973.
Training Required
>1 week training
Training Available
Recommend 3-day training course for persons unfamiliar with Chemstation software.
Manuals Available
Operator’s manual
Warranties
1 year
IATA Shipping Restrictions
Yes; Compressed gas cylinders must be shipped ground only according to IATA restrictions.
Please enter contact info.
Email Address: 
Company: 
Name: 
Title: