CT-1128
The CT-1128 is a lightweight, ruggedized, field serviceable GC/MS designed for environmental assessments, arms control and nonproliferation activities, and detection and identification of hazardous substances. The MS is a repackaged Agilent 5973N with two turbopumps for faster pump down.
Availability: DISCONTINUED
Neal Werner
+1 727 547 0600
7887 Bryan Dairy Road, Suite 100
Largo, Florida 33777-1452
|
|
www.contech.com
| Manufacturer |
| Constellation Technology Corporation |
| Product Name |
| CT-1128 |
| Detection Category |
| Chemical |
| Detection Principle |
| Spectrometry; |
| Detection Method |
| Quadrupole Mass Spectrometry; |
| Application |
| Gas Analysis; Laboratory Analysis; |
| Equipment Type |
| Instrument |
| Product Synopsis |
| The CT-1128 is a lightweight, ruggedized, field serviceable GC/MS designed for environmental assessments, arms control and nonproliferation activities, and detection and identification of hazardous substances. The MS is a repackaged Agilent 5973N with two turbopumps for faster pump down. |
| Availability |
| DISCONTINUED |
| Technology Readiness Level (TRL) |
| 9 |
| User Feedback Sources |
| Indifferent/No user feedback |
| T&E Reports and Standards |
| Reports: CT1128_Final_128.pdf; CT1128_Final_128.pdf; CT1128_Final_128.pdf;
|
| Dimensions |
| 15 x 23 x 15 in (38 x 28.4 x 38 cm) |
| Weight |
| >50 lbs.; 75 lbs. (34 kg) |
| Power Requirements |
| AC/DC Line Power; 120 V AC, 5 A peak power |
| Transportability |
| Mounted |
| Ruggedness |
| Vibration/Shock; Shock-mounted for use in vehicle. |
| Operating Conditions |
| 32 to 110°C |
| Consumables |
| >$1,000/year; Typical GC supplies (ferrules, liners, etc.). Uses hydrogen carrier gas supply from internal canister; 160 hr operating time per canister. |
| Calibration Schedule |
| Yes |
| Suggested Routine Maintenance |
| Typical for GC/MS |
| Communications Interface |
| Laptop |
| Unit Cost |
| $139,500.00 |
| Maintenance Cost |
| $5 to $10K per year |
| Tested For Chemical Agents |
| Yes. Tested at SBCCOM. |
| Tested For ITF-25 TIC/TIMs |
| Yes |
| Tested For Explosives |
| Yes |
| Other Chemical Targets |
| Yes |
| Sample Introduction |
| Liquid; Solid |
| Sensitivity/Detection Limits |
| %; mg |
| Known Interferents/Inhibitors |
| GC/MS is highly selective for volatile and semivolatile organic compounds when used by a skilled operator. |
| Start Up Time (From Cold Start To Sample Ready) |
| 60 to 90 minutes from pumpdown to operational ready for analysis. |
| Response Time (Sample Application To Output) |
| >15 minutes; 5 to 30 minutes (GC runtime) |
| Alarm Capability |
| Audible and visual alarms |
| Software Control |
| Laptop computer with Agilent Chemstation software. NIST/AMSID compatible. |
| Other Operational Parameters |
| Same as Agilent 5973. |
| Training Required |
| >1 week training |
| Training Available |
| Recommend 3-day training course for persons unfamiliar with Chemstation software. |
| Manuals Available |
| Operator’s manual |
| Warranties |
| 1 year |
| IATA Shipping Restrictions |
| Yes; Compressed gas cylinders must be shipped ground only according to IATA restrictions. |