Mx3000P Realtime PCR

The device is a real-time PCR instrument capable of four dye multiplexing. The Mx3000P's single source, single detector optic system is unique to this instrument class and is optimized for reliability and sensitivity without the need for a reference dye. Quartz tungsten halogen lamp with excitation range between 350 to 750 nm. The detection source is a single photomultiplier tube (PMT) with emission ranges from 350 to 700 nm. A computer can be provided with each Mx3005P system. The sample format includes 200 µL 96 well format, plates, 8-strip tubes, individual assay tubes. An Mx3000P is also available, with 4 filters rather than 5. Open platform design supports all fluorescent dyes and chemistries.

Notify me on updates of this product 

Availability: DISCONTINUED

Heidi Kijenski
Clinical Marketing Director
Heidi.Kijenski@agilent.com
USA
www.agilent.com
Manufacturer
Agilent Technologies
Product Name
Mx3000P Realtime PCR
Detection Category
Biological
Detection Principle
Molecular; Molecular Spectroscopy;
Detection Method
Polymerase Chain Reaction; Fluorescence ;
Application
Detection; Laboratory Analysis;
Equipment Type
Instrument
Product Synopsis
The device is a real-time PCR instrument capable of four dye multiplexing. The Mx3000P's single source, single detector optic system is unique to this instrument class and is optimized for reliability and sensitivity without the need for a reference dye. Quartz tungsten halogen lamp with excitation range between 350 to 750 nm. The detection source is a single photomultiplier tube (PMT) with emission ranges from 350 to 700 nm. A computer can be provided with each Mx3005P system. The sample format includes 200 µL 96 well format, plates, 8-strip tubes, individual assay tubes. An Mx3000P is also available, with 4 filters rather than 5. Open platform design supports all fluorescent dyes and chemistries.
Availability
DISCONTINUED
Technology Readiness Level (TRL)
9
User Feedback Sources
Indifferent/No user feedback
Dimensions
14 x 29 x 18 in (35.6 x 50.8 x 45.7 cm)
Weight
<50 lbs.; 45 lbs. (20.4 kg)
Power Requirements
AC/DC Line Power; 100 to 120 V AC, 50/60 Hz, 10 A 200 to 230 V AC, 50/60 Hz, 5 A
Noise Produced
None
Transportability
Person-Portable
Ruggedness
None
Operating Conditions
18 to 35°C (64 to 95°F)
Consumables
None
Solvents/Reagents
None
Suggested Routine Maintenance
Bulb replacement for fluorophore excitation
Data Analysis Support Equipment
Computer
Communications Interface
RS-232 to computer interface
Unit Cost
$29,315.00
Tested For Biological Agents
RNA/DNA
Other Biological Targets
Amenable based on technology
Adaptability
Yes; user can add new targets in the field
Sample Introduction
Liquid
Sensitivity/Detection Limits
Assay dependent (10 orders of magnitude dynamic range)
False Positive Rates
<1%
False Negative Rates
<1%
Start Up Time (From Cold Start To Sample Ready)
15 minutes
Response Time (Sample Application To Output)
<15 minutes
Other Operational Parameters
Recovery Time: Depends on fluorescent probe selection and data transfer settings; some data transfers take 15 minutes.
Training Required
<1 day training
IATA Shipping Restrictions
No
Please enter contact info.
Email Address: 
Company: 
Name: 
Title: