Sabre 4000

DISCONTINUED: Please see the Sabre 5000. The SABRE 4000 was an upgraded version of the Sabre 2000. It is programmed to detect and identify over 40 of threat substances in approximately 15 seconds. The vendor claims a cold start time of 10 minutes. The unit weighs approximately 7 lbs, including the 4-hr battery, and is outfitted with added TIC detection capability and a color 3.5 in TFT display.

Notify me on updates of this product 

Availability: DISCONTINUED

+1 410 612 2625
2202 Lakeside Boulevard
Edgewood, MD 21040
USA
www.smithsdetection.com
Manufacturer
Smiths Detection, Inc.
Product Name
Sabre 4000
Detection Category
Chemical;
Detection Principle
Spectrometry;
Detection Method
Ion Mobility Spectrometry (IMS);
Application
Emergency Response; Gas Analysis;
Equipment Type
Instrument
Product Synopsis
DISCONTINUED: Please see the Sabre 5000. The SABRE 4000 was an upgraded version of the Sabre 2000. It is programmed to detect and identify over 40 of threat substances in approximately 15 seconds. The vendor claims a cold start time of 10 minutes. The unit weighs approximately 7 lbs, including the 4-hr battery, and is outfitted with added TIC detection capability and a color 3.5 in TFT display.
Availability
DISCONTINUED
Technology Readiness Level (TRL)
9
Dimensions
14.5 x 4 x 4.5 in (36.3 x 11 x 13 cm)
Weight
7 lbs, including 4-hr battery
Power Requirements
Lithium battery (2-hr or 4-hr lifetime), 120/240 V AC power supply. 4-hr battery exceeds DOT regulation for transport on passenger aircraft.
Transportability
Person-Portable
Durability
Fair/good
Operating Conditions
-20 to 45ºC (14 to 113°F)
Consumables
Verification standards, air purification cartridge (APC), membrane and associated gaskets. APC is three times larger than Sabre 2000.
Calibration Schedule
Yes.
Suggested Routine Maintenance
Yes. Replace air purifier cartridge, replace membrane. System easily overloads when exposed to high analyte concentrations.
Available Accessories
Spares, toolkit, spare batteries, carrying case
Communications Interface
Direct serial connection to PC or laptop
Expected Operational Life
4 hour battery life
Shelf Life
Indefinite
Unit Cost
$25,000.00
Maintenance Cost
Estimated $2 to $3K/year.
Tested For Chemical Agents
GA, GB, GD, GF, VX, RVX.
Tested For ITF-25 TIC/TIMs
Yes. Unit produces a response for most organic compounds.
Tested For Explosives
PETN, TNT, nitroglycerin, RDX
Tested For Narcotics
Yes
Other Chemical Targets
Narcotics
Customize For New Targets
Yes. Customization should be performed by a trained technician using reference standards of the target chemical.
Sample Introduction
Vapor/Gas; Solid
Sensitivity/Detection Limits
ng; ppb; ppt
Known Interferents/Inhibitors
Testing at ECBC with the Sabre 2000 showed that the unit is susceptible to false positives in the presence of several tested interferents, including AFFF, bleach, fuels, cleaning products, and smokes. See the ECBC Report (August 2001) for details. The Sabre is also susceptible to overloading by high concentrations of target chemicals or interferents and must be baked out for 1 to 8 hr, depending on the degree of contamination, before it can be returned to service.
Start Up Time (From Cold Start To Sample Ready)
15 to 30 minutes
Response Time (Sample Application To Output)
15 to 30 seconds, depending on operating mode
Alarm Capability
Audible and visual, detected substance is indicated on panel display
Software Control
Unit is controlled by an on-board computer with color LCD display and four function buttons. Up to 200 analysis results can be stored directly on the unit and can be downloaded to a PC via RS-232 or USB port.
Other Operational Parameters
The user can select among four sets of operating parameters, called Control Parameter (CP) files: Explosives, Narcotics, CW-POS, and CW-NEG. Only one CP may be used at a time. Switching between CPs is straightforward but there may be up to a 1 to 5 minute delay for internal temperatures to stabilize to new settings.
Operator Skills Required
Non-technical background
Training Available
Yes
Manuals Available
Operator’s manual
Warranties
1 year
ITAR Export Regulations
U.S. Nuclear Regulatory Agency regulations due to 63Nickel source.
IATA Shipping Restrictions
U.S. Nuclear Regulatory Agency regulations due to 63Nickel source.